CHAPTER 1 MAINTENANCE TESTS
1–18 www.zoll.com 9650-0309-01 Rev. E
12.0 Real Time Clock Test
Equipment Needed
Power Supply 12V 10A,
Battery Current Test Fixture (P/N 9100-0690-TF) or AED PRO Disposable Battery (P/N
1008-002003-01),
AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF).
Setup:
Connect AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF) to the unit.
Place the Battery Current Test Fixture (P/N 9100-0690-TF) in the battery well.
Set the Power Supply to 12.00V.
Turn on the Power Supply and turn the unit off.
Note: To set the current date and time in the AED pro, make sure you are in
Non-Rescue Mode.
Test Action Expected Result(s) Pass Fail
12.1
Press and hold the On/Off
button for at least 7 seconds,
then release it.
Verify the unit issues a voice
and text prompt UNIT OK,
12.2
Press SET TIME softkey. Verify the unit prompts you to
set the Year, Month, Day Hour,
Minute, and Second. To advance
to each subsequent screen, press
softkey.