Test Equipment
9650-0309-01 Rev. E ZOLL AED Pro Service Manual 1–19
12.3
SET YEAR: Press the + Softkey
to the desired year, using
numbers 2000 through 2037.
Press the softkey to advance
to the next screen.
SET MONTH: Press the +
Softkey to the desired month,
using numbers 01 through 12.
SET DAY: Press the + Softkey
to the desired day, using
numbers 01 through 31.
SET HOUR: The AED Pro has a
24-hour clock. Press the +
Softkey to the desired hour,
using numbers 01 through 24.
SET MINUTE: Press the +
Softkey to the desired minute,
using numbers 00 through 59.
SET SECOND: Press the +
Softkey to the desired second,
using numbers 00 through 59.
Press the softkey one more
time.
Note: You cannot go backwards
when scrolling through the
numbers.
The unit then displays the
selected date and time.
12.4
Press the left softkey (Enter). The Text message SAVING
DATA appears.