EasyManua.ls Logo

ZOLL aed pro - Page 20

ZOLL aed pro
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CHAPTER 1 MAINTENANCE TESTS
1–8 www.zoll.com 9650-0309-01 Rev. E
.
5.7
Press VFIB on the Defibrillator
Analyzer.
After 2 minutes (default setting),
Verify the unit prompts STOP
CPR, followed by DON’T
TOUCH PATIENT
ANALYZING twice, followed
by SHOCK ADVISED,
DON’T TOUCH PATIENT,
PRESS FLASHING SHOCK
BUTTON.
5.8
Press Flashing Shock button Verify that unit prompts
SHOCK DELIVERED.
5.9
2
nd
shock energy test (default
value 150 joules)
Verify the Energy delivered into
analyzer is 122 joules to 166
joules.
5.10
Verify the unit prompts START
CPR, followed by CONTINUE
CPR.
5.11
Press VFIB on the Defibrillator
Analyzer.
After 2 minutes (default setting),
Verify the unit prompts STOP
CPR, followed by DON’T
TOUCH PATIENT
ANALYZING twice, followed
by SHOCK ADVISED,
DON’T TOUCH PATIENT,
PRESS FLASHING SHOCK
BUTTON.
5.12
Press Flashing Shock button Verify that unit prompts
SHOCK DELIVERED.
5.13
3
rd
shock energy test (default
value 200 joules)
Verify the Energy delivered into
analyzer is 155 joules to 211
joules.
5.14
Press and hold the On/Off
button to power off the unit.
Verify unit displays UNIT
POWERING OFF message.