CHAPTER 1 MAINTENANCE TESTS
1–12 www.zoll.com 9650-0309-01 Rev. E
8.6
Verify the ready beep stays
steady for 3 seconds
approximately and voice
prompts again DON’T TOUCH
PATIENT, PRESS
FLASHING SHOCK
BUTTON, followed by 3 beeps,
and then NO SHOCK
DELIVERED prompt.
8.7
Press and hold the On/Off
button to power off the unit.
Unit powers off.