CHAPTER 1 MAINTENANCE TESTS
1–14 www.zoll.com 9650-0309-01 Rev. E
9.5
Remove the Decade Box.
Select GICAL from ‘Command
1’ window and click SEND.
The response will be SILLO,
SILMED, SILHI. Compare as
followed:
56 < SILLO< 176 counts (10
Ohms),
134 < SILMD< 321 counts (35
Ohms),
972 < SILHI< 2129 counts (300
Ohms).
9.6
Select SFCAL from ‘Command
1’ window and click SEND.
The response will be SFCAL,
OK.
Note: the Calibration values will
go back to default conditions if
not saved. Save calibration
values. both ECG offset and
Impedance must be calibrated.