266CRH / 266CRT, 266CSH / 266CST MULTIVARIABLE PRESSURE TRANSMITTER | OI/266CXX/MODBUS-EN REV. A 53
MV Setup - closed Tank
MV-Calculation No Calculation
Flow
Level
Level Calculation
Level Unit Steam Drum Level
Open Tank
Closed Tank
Closed Tank, 2 Remote
Seals
Lower Level Dimension b from Figure 37
Upper Level Dimension d from Figure 37
Distance process Connection Dimension e from Figure 37
Transm. mount. Pos. Dimension a from Figure 37
Temperature Unit
Min. Temperature Minimum measuring medium temperature
Max. Temperature Maximum measuring medium temperature
Density Unit
Min. Temp. Density Density of the measuring medium at minimum temperature
Max. Temp. Density Density of the measuring medium at maximum temperature
Temperature Gas Density Reference temperature for specification of the gas density, for example 0 C for ಫstd.
conditionsಬ
Static Pressure Unit
Pressure Gas Density Reference pressure for specification of gas density, for example 1.013 bar for ಫstandard
conditionsಬ
Gas Density Gas density for the reference conditions specified above
Tube Temperature PT100
Sensor Temperature
Fixed Value
Value Condensate
Temperature
Restore Last Parameter