266CRH / 266CRT, 266CSH / 266CST MULTIVARIABLE PRESSURE TRANSMITTER  |  OI/266CXX/MODBUS-EN REV. A  39 
 
 
… 7  Commissioning 
 
Parameterization of the device 
 
The transmitter is delivered preconfigured according to the 
information provided when placing the order. 
However, should a change to the configuration be necessary 
(because measuring point data has changed since the original 
plans were drawn up, for example), the following options are 
available: 
•  Menu-led configuration of the transmitter with the 
integral LCD display 
•  Configuration using a PC / Laptop with graphical user 
interface (DTM). How to use these tools to make the 
configuration settings is described in the corresponding 
related documentation. 
 
 
Factory settings 
 
The transmitters calibrated in the factory to the measuring 
range specified by the customer. The calibrated measuring 
range and the measuring point tag are specified on an additional 
labeling plate. 
If nothing is specified by the customer in this regard, the 
transmitter will be delivered with a standard configuration, that 
contains the following parameters (among others). 
 
Parameter Factory setting 
Measuring range start (LRV)  Zero 
Measuring range end (URV)  Upper measuring range limit (URL) 
Transmission function for the output  Mass flow 
 
Damping 1 second 
Presentation of the optional LCD 
display 
Process value PV (1-place) and bar 
diagram of the output signal 
 
Each of the parameters listed here can be easily set via the 
optional LCD display with operating menu, or the Device Type 
Manager (DTM). 
 
 
 
 
 
Configuration without integrated LCD 
display 
Note 
The configuration possibilities described below are only possible 
for the models 266Cxx with flow calculation or without 
calculation function. 
 
The correction of the mounting position and reset of bias are set 
directly on the transmitter via the operating buttons. These 
operating buttons are arranged under the rating plate. 
In order to operate the device locally, unscrew the fixing screws 
of the rating plate, and swing the name plate clockwise to the 
side. 
 
NOTICE 
Material damage due to a magnetic field! 
The use of magnetic screwdrivers results in damage of 
components. 
•  Do not use a magnetic screwdriver to operate the buttons. 
 
The transmitter has been calibrated by the manufacturer based 
on the order information. The set measuring range start and 
measuring range end are specified on the name plate. 
 
For the connection, ensure that liquid residues (for gaseous test 
materials) or air bubbles (for liquid test materials) are not in the 
impulse line; because that can cause measurement errors in the 
test. 
 
It is recommended to set the damping to the value ‘zero’. 
For correction of mounting position and static pressure 
influence see . 
A reset to factory setting of differential pressure can be done 
with operating button ‘S’.