56 266CRH / 266CRT, 266CSH / 266CST MULTIVARIABLE PRESSURE TRANSMITTER | OI/266CXX/MODBUS-EN REV. A
… 8 Operation
… Overview of parameters on the configuration level
MV Setup - 2 Remote Seals
MV-Calculation No Calculation
Flow
Level
Level Calculation
Level Unit Steam Drum Level
Open Tank
Closed Tank
Closed Tank, 2 Remote
Seals
Lower Level Dimension b from Figure 39
Upper Level Dimension d from Figure 39
Distance process Connection Dimension e from Figure 39
Transm. mount. Pos. Dimension a from Figure 39
Temperature Unit
Min. Temperature Minimum measuring medium temperature
Max. Temperature Maximum measuring medium temperature
Density Unit
Min. Temp. Density Density of the measuring medium at minimum temperature
Max. Temp. Density Density of the measuring medium at maximum temperature
Temperature Gas Density Reference temperature for specification of the gas density, for example 0 C for std.
conditions
Static Pressure Unit
Pressure Gas Density Reference pressure for specification of gas density, for example 1.013 bar for std.
conditions
Gas Density Gas density for the reference conditions specified above
Capillary Temperatur PT100
Sensor Temperature
Fixed Value
Value Capillary Temperature
Fill Fluid Halocarbon 4.2, Galden G5,
DC 200, AN 140, Syltherm,
Silicon M5, Neobee M20,
Glycerin Water, Marcol 122,
Silicon DC704
Restore Last Parameter