EasyManuals Logo

ABB AO2000 Series User Manual

ABB AO2000 Series
362 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
106 AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Connection of the sample gas line
1
Sample gas line (heated or unheated, PTFE or stainless steel tube with
inside/outside diameter 4/6 mm)
Connection either with
2
O-ring 6.02 x 2.62
3
Socket
4
O-ring 12.42 x 1.78
or with
5
Male fitting G¼”
or with
6
Socket
7
Tapered ring
8
Compression fitting
9
Exhaust gas tube for 19-inch housing
10
Exhaust gas tube for wall-mount housing
Maximum length of the sample gas line
The maximum length of the unheated sample gas line (inside diameter
4 mm) is 50 m.
Provide for sample gas line purging
Install a shut-off valve in the sample gas line (definitely recommended with
pressurized sample gas), in order to provide a means for purging the gas
line system by feeding in an inert gas, e.g. nitrogen, from the gas sampling
point.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB AO2000 Series and is the answer not in the manual?

ABB AO2000 Series Specifications

General IconGeneral
BrandABB
ModelAO2000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals