EasyManua.ls Logo

ABB AO2000 Series - Standard Configuration

ABB AO2000 Series
362 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B 181
Standard configuration
Standard configuration
Various applications are factory-configured. These standard configurations
are based on
standard input/output pin configuration and
the available sample components
Some factory-configured applications require field linking of additional
function blocks.
Example: Limit value monitoring
The limit value monitoring application consists of a factory configured link
between
Component measurement value, Hold, Limit monitor and Digital output
function blocks.
Grenzwert-
waechter
‘’
Komp.-
Messw.
‘’
Digital-
ausgang
‘’
Halteglied
‘’
Auto-
kalibrierung
‘’
Example: Measuring range switch-over/feedback signal
The measurement range switch-over/feedback application consists of a
factory-configured linkage between
a Measuring range switch-over function block with several Digital input
function blocks and Component range function block, as well as
a Range feedback function block with the same Component measuring
range function block and several Digital output function blocks.
Messber.-
um-
schaltung
‘’
Messber.-
rueck-
meldung
‘’
Digital-
ausgang
‘’
Komp.-
Messber.
‘’
Digital-
eingang
‘’
Digital-
eingang
‘’
Digital-
ausgang
‘’

Table of Contents