162 AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Limit value monitor parameterization
Menu path
MENU Configure Component-specific Limit Values
Select Limit Monitor ...
Range selection
All available limit value monitors are shown.
Procedure
Parameter Explanation Action
Direction
< = Alarm on underflow of threshold value
> = Alarm on exceeding threshold value
Select
Threshold
in physical units Set
Hysteresis
in physical units Set
Standard configuration
As a rule, limit value monitoring for those components to be measured by
the gas analyzer is factory-set. This requires that there be enough digital
outputs on the I/O modules to handle the number of sample components.
Note: Limit value monitors are factory-set or user-configured Limit Monitor
function blocks. The Technical Information "Function Blocks – Descriptions
and configuration" contains complete information on the individual function
blocks.