AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B 341
Pneumatic module: Troubleshooting
Flow error
Condensation in the flowmeter
Disconnect gas analyzer from the gas conditioning system.
Dry the flow meter by heating it and blowing it with compressed air.
Check the operation of the upstream sample gas cooler.
Inadequate gas supply
Connect the flow meter, ball chamber or pressure gauge directly to the
gas supply pump and measure the pressure or vacuum.
Check the pump and change the membrane if necessary.
Check the disposable filter and replace (see page 314) if necessary.
Check the solenoid valve(s) and replace if necessary.