334 AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Caldos25, Caldos27, Magnos206, Magnos28, Magnos27:
Troubleshooting
Temperature problem
Connecting leads of the temperature sensor or the heater discon-
nected
Check the connecting leads and connectors.
Check the line seating in the insulated jackets.
Defective thermal link
Check thermal link for continuity and replace if necessary.
Leaking thermostat or open purge gas connections
Check the seal integrity between the thermostat chamber and the
mounting flange; tighten mounting bolts or replace O-rings as needed.
Seal the analyzer purge gas inlet and outlet with sealing connectors.
Check the purge gas and reduce as needed: in operation max. 20 l/h,
positive pressure p
e
= 2 to 4 hPa.
Unstable readings
Gas path leakage
Check the seal integrity (see page 268) of the analyzer module gas paths
and (if applicable) of the lines to the pneumatics module.