EasyManua.ls Logo

ABB AO2000 Series - Configuring Bus IOs

ABB AO2000 Series
362 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B 197
Configuring Bus I/Os
Menu path
MENU Configure System Network BUS IO
Number of Bus I/Os
Changing the number of Bus I/Os will affect the Modbus address range,
Profibus map and Ethernet linking.
Reducing the number of Bus I/Os may cause transmission errors if the set-
tings in the communication partners are not matched. This may also cause
damage in the function block applications.
Parameters
Parameter Function Read Write Example
BUS AI
Bus analog
inputs
x x for analog value input into the function block
application
BUS AO
Bus analog
outputs
x for analog value output from the function
block application
BUS DI
Bus digital
inputs
x x for control of functions such as auto calibra-
tion, measurement range control after func-
tion block configuration
BUS DO
Bus digital
outputs
x for display of functions linked by function
block configuration, e.g. alarm signaling

Table of Contents

Related product manuals