EasyManuals Logo

ABB AO2000 Series User Manual

ABB AO2000 Series
362 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #123 background imageLoading...
Page #123 background image
AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B 123
24-VDC connection
1 Connect the supplied connection cable with receptacle to the 24-V DC
connection on the analyzer module.
2 Connect the wires on the free end of the cable to the power supply filter
-Z01 in the central unit or to the external power supply.
Extending the 24-VDC connection cable
The 24-VDC connection cable has a wire section of 0.5 mm
2
; it is limited to a
length of 5 meters (16.4 feet).
Note the following conditions if the connection cable has to be extended:
• The extension cable must have a wire section of at least 2.5 mm
2
.
• The extension cable should be no more than 30 meters (100 ft) long.
• The extension cable should be connected as close as possible to the
receptacle of the supplied 24-VDC connection cable, i.e. the supplied
24-VDC connection cable should be made as short as possible.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB AO2000 Series and is the answer not in the manual?

ABB AO2000 Series Specifications

General IconGeneral
BrandABB
ModelAO2000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals