172 AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Changing a component
The (maximum four) components, which are displayed in the component list
(see page 165), can be changed. Either the parameters of a component can
be changed or one component can be replaced by another.
Menu path
MENU Configure Component-specific Active Component
Select component and confirm with ENTER.
Change parameter of a component
NOTES
In the case of a system component (Type S) only the parameter "Unit" can be
changed; in the case of a user component (Type B) all the parameters can be
changed with the exception of the parameter "Type".
The parameters of the active component (Type SA or BA) and the compo-
nent used for calibration (Type SK or BK) cannot be changed.
1 Press softkey COMP. CONFIG. The "Configure Component" menu is
displayed with the component list.
2 In the component list select the component with parameters which are
to be changed.
3 Press softkey CHANGE COMP. Enter password (level 3), as applicable. The
parameters of the component are displayed.