AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B 331
No. Text S A
W
F Reaction/Remark
Miscellaneous Messages
1000 This function block has an er-
ror:
x x Is supplemented with a reference to the function block
type.
1001 Condensate penetration. Overall message for guiding reaction to condensate
penetration; not logged
1002 Flow rate too high at this point! x x Currently not used
1003 Flow rate too low at this point! x x Currently not used
System Cooler
1100 Cooler temperature too high. x x Sample gas feed module pump is automatically turned
off. Check the system cooler and sample gas prepara-
tion system.
1101 Cooler temperature too low. x x
1102 Condensate penetration in
cooler.
x x
1103 Flow rate too low in cooler. x Check the system cooler and sample gas preparation
system.
1104 Cooler condensate level too
high.
x
1105 Cooler condensate level too
high.
x Empty the condensate bottle.
1106 Cooler reagent level too low. x Fill the reagent container.