102 AWT420 | UNIVERSAL 4-WIRE, DUAL-INPUT TRANSMITTER | OI/AWT420-EN REV. B
Appendix C Dual 2-electrode conductivity calculations
Monitoring on AVT systems
For low conductivity feedwater applications, all volatile
chemical treatment (AVT) is often applied.
Where cation resin columns are used to remove the effects on
the conductivity measurement of volatile ammonia and
hydrazine chemical treatment, it is common practice to
measure both before- and after-cation conductivity. The
sensitivity of the conductivity measurement to chemical
treatments is increased by passing the sample through the
cation column.
If it is known that a sample contains only one impurity (e.g.,
ammonia), the conductivity measurement now becomes an
indication of the concentration of that impurity and it is then
possible to calculate the pH of the sample from the
concentration data. The result is referred to as 'inferred pH'.
The maximum after-cation conductivity value is programmable

-1
dependent on local
conditions. After-cation values above this level generate an
AFTER CAT. HIGH error message and before-cation values

-1
generate a BEFORE CAT. HIGH error
message. The inferred pH range is 7 to 10 pH, values above
10pH generate an Infr. pH invalid error message.
Refer to page 83 for message descriptions.
The inferred pH feature can be used on AVT systems only in
the following circumstances:
1 On steam raising plant.
2 For boiler chemical treatment such as ammonia and/or
hydrazine. In this instance, A: Temp. Comp. must be set to
NH3 and B: Temp. Comp. must be set to ACID.
3 Where the after-cation conductivity value is insignificant
compared to the before-cation value.
NOTICE
Inferred pH measurement on AVT systems is inappropriate
to chemical treatments such as sodium phosphate, sodium
hydroxide and morpholine.
Monitoring on AVT systems with impurities

pH on AVT systems where there are low concentrations of ionic
impurities present in addition to the volatile alkaline agent (e.g.
sodium chloride + ammonia). In this case, the exchange of
ammonium and sodium ions within the cation column releases
water and hydrochloric acid. The sodium chloride impurity
produces a conductivity after the column that is higher than
the conductivity before.
The dual input analyzer, when used to monitor before- and
after-cation conductivities, compensates for this increase and
calculates the inferred pH of the incoming sample. The user-
configurable, after-cation conductivity alarm can be used to
detect unacceptably high levels of impurities in the sample and
inform the user of the validity of the inferred pH value.
The calculated inferred pH is proportional to:

Where:
  
  
  

-1
  
and AC readings
The maximum after-cation conductivity value is programmable

-1
dependent on local
conditions. After-cation values above this level generate an
AFTER CAT. HIGH error message and before-cation values

-1
generate a BEFORE CAT. HIGH error
message. The inferred pH range is 7 to 10 pH, values above 10
pH generate an Infr. pH invalid error message. Refer to page
83 for a description of error messages.
The inferred pH feature can be used on AVT systems with
impurities only in the following circumstances:
1 On steam raising plant.
2 For boiler chemical treatment such as ammonia and/or
hydrazine. In this instance, A: Temp. Comp. must be set to
NH3 and B: Temp. Comp. must be set to ACID.
3 Where the after-cation conductivity value is less than

-1
.
NOTICE
Inferred pH measurement on AVT systems is inappropriate
to chemical treatments such as sodium phosphate, sodium
hydroxide and morpholine.