64 AWT420 | UNIVERSAL 4-WIRE, DUAL-INPUT TRANSMITTER | OI/AWT420-EN REV. B
…15 Calibration procedures
… pH/Redox/ORP calibration
1-Point automatic calibration
Note. Before starting the calibration ensure the pH buffers
are set to the correct values.
1 At the Calibrate level, press the key:
The Calibrate menu is displayed:
2 Use the / keys to select the sensor to be
calibrated, and press the key to confirm selection.
The pH calibration menu is displayed:
3 Use the / keys to select sensor calibration and
press the key to confirm selection.
The calibration type is displayed:
4 Use the key to edit the calibration type.
Use the / keys to select the required calibration
type and press the key to confirm selection.
Press the key to proceed to the next step.
5 Place the sensor into buffer 1 and press the
key to perform the calibration. The calibration
process screen is displayed.
On completion the result screen is displayed.
If the calibration passes, the slope and offset values are
displayed.
If the calibration fails, the failure reason is displayed.
Note. The calibration can be canceled at any time during
the process by pressing the Abort key.
Calibrate
Back
Select
S1 : pH/Redox (ORP)
S2 : pH/Redox (ORP)
pH Buffers
Hold Outputs
Calibrate
Back
Select
Sensor Cal
Temperature Cal
Edit Cal
Sample Collection
Collection Complete
Restore Cal Defaults
Sensor Cal
Next
Edit
Calibration Type
1-Pt Auto
Calibrate
Abort
Continue
Immerse in Buffer 1
ABB
4.0 pH
Calibrate
Abort
pH 4.00 pH
Tmp 25.0 °C
Settling-Please Wait