64 AWT420 |  UNIVERSAL 4-WIRE, DUAL-INPUT TRANSMITTER  |  OI/AWT420-EN REV. B
…15   Calibration procedures
… pH/Redox/ORP calibration
1-Point automatic calibration
Note. Before starting the calibration ensure the pH buffers  
are set to the correct values.
1  At the Calibrate level, press the   key:
The Calibrate menu is displayed:
2  Use the  /  keys to select the sensor to be  
calibrated, and press the   key to confirm selection.
The pH calibration menu is displayed:
3  Use the  /  keys to select sensor calibration and  
press the   key to confirm selection.
The calibration type is displayed:
4  Use the   key to edit the calibration type. 
Use the  /  keys to select the required calibration  
type and press the   key to confirm selection.
Press the   key to proceed to the next step. 
 
5  Place the sensor into buffer 1 and press the   
key to perform the calibration. The calibration  
process screen is displayed.  
 
On completion the result screen is displayed.
•  If the calibration passes, the slope and offset values are 
displayed. 
•  If the calibration fails, the failure reason is displayed. 
Note. The calibration can be canceled at any time during  
the process by pressing the Abort   key.
Calibrate
Back
Select
S1 : pH/Redox (ORP)
S2 : pH/Redox (ORP)
pH Buffers
Hold Outputs
Calibrate
Back
Select
Sensor Cal
Temperature Cal
Edit Cal
Sample Collection
Collection Complete
Restore Cal Defaults
Sensor Cal
Next
Edit
Calibration Type
          1-Pt Auto
Calibrate
Abort
Continue
Immerse in Buffer 1
ABB 
4.0 pH
Calibrate
Abort
pH    4.00 pH
Tmp   25.0 °C
Settling-Please Wait