62 AWT4 20 |  UNIVERSAL 4-WIRE, DUAL-INPUT TRANSMITTER  |  OI/AWT420-EN REV. B
…15   Calibration procedures
…pH/Redox/ORP calibration
2-Point calibration
1  At the Calibrate level, press the   key:
The Calibrate menu is displayed:
2  Use the  /  keys to select the sensor to be  
calibrated, and press the   key to confirm selection.
The pH calibration menu is displayed:
3  Use the  /  keys to select sensor calibration and  
press the   key to confirm selection. 
The calibration type is displayed:
4  Use the   key to edit the calibration type. 
Use the  /  keys to select the required calibration  
type and press the   key to confirm selection.
Press the   key to proceed to the next step. 
 
The buffer temperature is displayed:
5  Use the   key to edit the buffer temperature. 
Use the  /  keys to set the temperature and press  
the   key to confirm changes.
Press the   key to proceed to the next step. 
 
The first buffer value is displayed:
Calibrate
Back
Select
S1 : pH/Redox (ORP)
S2 : pH/Redox (ORP)
pH Buffers
Hold Outputs
Calibrate
Back
Select
Sensor Cal
Temperature Cal
Edit Cal
Sample Collection
Collection Complete
Restore Cal Defaults
Sensor Cal
Next
Edit
Calibration Type
          2-Pt Manual
1
2
3
Sensor Cal
Next
Edit
Buffer Temperature
               25.0 °C
1
2
3
Sensor Cal
Next
Edit
Low Buffer Value
               4.0 pH