AWT420 | UNIVERSAL 4-WIRE, DUAL-INPUT TRANSMITTER | OI/AWT420-EN REV. B
69
Temperature calibration
1 At the Calibrate level, press the key:
The Calibrate menu is displayed:
2 Use the / keys to select the sensor to be calibrated,
and press the key to confirm selection.
The pH calibration menu is displayed:
3 Press the / keys to select Temperature Cal and press
the key to confirm selection.
The temperature calibration screen is displayed:
4 Wait for the displayed value to stabilize and press the key
to proceed to the next step.
The buffer temperature is displayed:
5 Use the / and keys to set the temperature and press
the key to confirm changes.
The temperature calibration process screen is displayed:
On completion the result screen is displayed.
If the calibration passes the slope and offset values are
displayed.
If the calibration fails the reason for failure is displayed.
Note. The calibration can be canceled at any time during the
process by pressing the Abort key ( ).
Calibrate
Back
Select
S1 : pH/Redox (ORP)
S2 : pH/Redox (ORP)
pH Buffers
Hold Outputs
Calibrate
Back
Select
Sensor Cal
Temperature Cal
Edit Cal
Sample Collection
Collection Complete
Restore Defaults
Temperature Cal
TMP 25.0 °C
Continue when Stable
Abort
Continue
Temperature Cal
TMP 25.0 °C
New 0025.0
Next
Continue
Temperature Cal
TMP 25.0 °C
Abort