EasyManua.ls Logo

ABB AWT420 - Connecting Ezlink Sensors; Long Cables

ABB AWT420
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20 AWT420 | UNIVERSAL 4-WIRE, DUAL-INPUT TRANSMITTER | OI/AWT420-EN REV. B
…5 Electrical installation
Connecting EZLink sensors
NOTICE

refer to sensor Operating instruction.
Referring to Figure 16:
1 Align the pins in sensor cable connector A with the holes in
EZLink connector B and push the connectors together.
2 Turn nut C clockwise to secure the connectors together.
The transmitter detects the type of sensor connected
automatically.
NOTICE
When installing sensor extension cables, ensure the male
end (end with label) of the cable is installed towards the
transmitter.
Long cables
If cables are longer than 30 m (94 ft), or they are outside, the
following cables must be screened or contained in conductive
conduit:
digital I/O
analog outputs
communication
Figure 16 Connecting the sensor EZLink connector(s)
A
B
C

Table of Contents