84 AWT420 | UNIVERSAL 4-WIRE, DUAL-INPUT TRANSMITTER | OI/AWT420-EN REV. B
…16 Troubleshooting
…Diagnostic messages
…AWT420 transmitter diagnostics
NAMUR
icon
Diagnostic
message
Cause Recovery action
Clean (n) in
progress
 
S(n):
Write Error
Error writing configuration to
sensor/sensor module.
Repeat previous configuration change.
If Sensor Write Error persists, cycle power to the transmitter.
Check the Sensor Setup and correct if necessary.
If Sensor Write Error persists, ensure that the sensor and the
transmitter are compatible by upgrading the software on both via
the bootloader.
Check the Sensor Setup.
If Sensor Write Error persists, contact local service organization.