158 EL3000 CONTINUOUS GAS ANALYZERS | OI/EL3000-EN REV. D
10 Configuration
Software updates
10. Confirm the update of software in the ‘EasyLine Query
dialog by clicking ‘Ok’.
The new software is loaded onto the gas analyzer.
11. After the update has been completed, the information ‘New
software successfully installed. Wait for the gas analyzer to
restart. After the restart, a new language can be loaded onto
the gas analyzer’ appears.
Close the window by clickingOk’.
The gas analyzer has been restarted (‘Booting’).
Note
The software update and restart of the gas analyzer can take
several minutes