IMMUNITY
test
IEC 60601
test level
Compliance
Level
Electromagnetic
environment – guidance
Conducted RF
IEC 61000-4-6
6 Vrms
150 kHz to
80 MHz
6 Vrms Portable RF communications
equipment (including peripherals
such as antenna cables and external
antennas) should be used no closer
than 30 cm (12 inches) to any part
of the System, including cables
specified by Abbott Diabetes Care.
Otherwise, degradation of the
performance of the System could
result.
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to
2.7 GHz
10 V/m
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
See table on
next page
Compliance to
the tested levels
The table below lists the immunity test levels at specific test frequencies for testing the effects
of some wireless communications equipment. The frequencies and services listed in the table are
representative examples in healthcare and in various locations where the System may be used.
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power (W)
Distance
(m)
IMMUNITY
TEST
LEVEL
(V/m)
385 380 – 390 TETRA 400 Pulse
modulation
b)
18 Hz
1.8 0.3 27
450 430 – 470 GMRS 460,
FRS 460
FM
c)
± 5 kHz
deviation
1 kHz sine
2 0.3 28
710
704 – 787 LTE Band 13,
17
Pulse
modulation
b)
217 Hz
0.2 0.3 9
745
780
810
800 – 960 GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2 0.3 28
870
930
129 130
ART43974_rev-A_manual.indd 129-130ART43974_rev-A_manual.indd 129-130 2/17/21 8:32 AM2/17/21 8:32 AM