EasyManua.ls Logo

Abbott FreeStyle Libre 2 - Control Solution Testing

Abbott FreeStyle Libre 2
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Display What To Do
If your blood ketone is high, you will see a
message on the screen. You can touch the
message button for more information.
High Ketone
HI
mmol
L
Ketone Test
If HI appears on the Reader, your ketone result is
higher than 8 mmol/L. You can touch the message
button for more information. Repeat the ketone
test with a new test strip. If you get a second
HI result, contact your health care professional
immediately.
1.7
mmol
L
Ketone Test
High Ketone
Control Solution Testing
You should do a control solution test when you are not sure of your
test strip results and want to check that your Reader and test strips are
working properly. You can do a control solution test with a blood glucose
or ketone test strip.
IMPORTANT:
Control solution results should fall within the control
solution range printed on the test strip instructions for use.
Do NOT use control solution past the expiry date. Discard
control solution 3 months after opening. Refer to control
solution instructions for use.
The control solution range is a target range for control
solution only, not for your blood glucose or ketone results.
The control solution test does not reect your blood glucose
or ketone levels.
Use only Optium glucose and ketone control solution.
Check that the LOT number printed on the test strip foil
packet and instructions for use match.
Replace the cap securely on the bottle immediately after use.
Do NOT add water or other liquid to the control solution.
Contact Customer Service for information on how to obtain
control solution.
63 64
ART43974_rev-A_manual.indd 63-64ART43974_rev-A_manual.indd 63-64 2/17/21 8:32 AM2/17/21 8:32 AM