U
nderstanding C ont rol Test
R
esults
20
Chapter 2: Control Testing ACCU-CHEK Customer Care Service Center 1-800-858-8072
The label on your test strip container shows the acceptable
ranges for low and high test solutions. Your con trol test
resul ts should be with in the listed range. Make sure you
co
m
pare the result to the range for the l evel of control
solution you used
(
low or high
)
. The low level control
solution bottle has a blue cap; the high level solution bottle
has a red cap.
W
hen a control test result is within the acceptable range,
your test strips and
m
eter are working properly.
R
ange
(
m
g/d
L
)
L
O
W
42–72
H
I
GH
290–328
Example
Example