7451A/7461A/7461P Digital Multimeter Operation Manual
6.4.2 Measurement Complete Signal Terminal (COMPLETE OUT)
6-10
Complete signal output timing
The typical value of the complete signal output timing under the following setting conditions is shown.
Function: DCV, Auto-zero: OFF, Auto-range: OFF
1. Single output
Time from the trigger signal input at the external trigger terminal to the complete signal output: Tc
Tc=Tin Sampling count + 2.5 msec
Tin: The longest time in the followings.
Sampling interval (SI)
Integration time (IT) + 30 sec
310 sec: Display OFF, Measurement data memory ON
850 sec: Display OFF, Measurement data memory OFF
1.29 msec: Display ON
2. Multi output
Period of the complete signal output: Tcp
Tcp=Tin+1.4 msec
Tin: The longest time in the followings.
Sampling interval (SI)
Integration time (IT) + 30 sec
310 sec: Display OFF, Measurement data memory ON
850 sec: Display OFF, Measurement data memory OFF
1.29 msec: Display ON