7451A/7461A/7461P Digital Multimeter Operation Manual
4.2.1 Measurement Functions
4-17
10. Temperature measurement (TEMP-2W, TEMP-3W, TEMP-4W)
Press .
The temperature is measured by using the 2W, 3W or 4W (formerly set method).
The measurement resolution varies depending on the sampling rate.
Pressing switches the sampling rate in turn.
Select the sampling rate according to the necessary resolution.
Table 4-14 Temperature Measurement
Resistance bulb Pt100 JPt100
Measurement range -200 C to +850 C -200 C to +649 C
Measurement current 1 mA
Table 4-15 Display Resolution
Sampling rate Integration time Measurement resolution
FAST1 0.02PLC 0.1 C
FAST2 0.1PLC 0.1 C
MED 1PLC 0.01 C
SLOW1 5PLC 0.001 C
SLOW2 10PLC 0.001 C
TEMP
RATE