7451A/7461A/7461P Digital Multimeter Operation Manual
6.7.3 ADC Command Reference
6-23
Measurement DCV-Bch
measurement mode
*  Valid only in the 
DCV-Bch func-
tion (F12)
SDV0 HI-COM measurement
SDV1 HI-LO measurement
SDV2 RATIO measurement
SDV? Reply: One of SDV0 to SDV2
Invalid function INH?n Reads the invalid state of the function.
n: 1 (DC voltage measurement) to 50 (Frequency 
measurement)
Reply 0: Valid state of the function selection
1: Invalid state of the function selection
Trigger *TRG Trigger command
Measurement data 
memory
ST0 Store: OFF
ST1 Store: ON
ST? Reply: ST0 or ST1
IRDn, m Readout range setting
n, m:  0 to 9999 (7451A/61A)
 0 to 19999 (7461P)
• Parameters cannot be omitted.
(0, 0)
IRO? Data reading
Reply: See Section 6.7.2, “Output Data Format.”
IRPO? Number of measurement results readings
Reply:  IRPOdddd (7451A/61A)
 IRPOddddd (7461P)
IRNO? Data range reading
Reply:  IRNOdddd, dddd (7451A/61A)
 IRNO
ddddd, 
ddddd (7461P)
No data: IRNO0000, -001 (7451A/61A)
 IRNO+00000, -00001 (7461P)
ICL Measurement data memory initialization
Trigger system Start INI Leaves the IDLE state.
Continue INIC0 CONTINUOUS OFF
INIC1 CONTINUOUS ON
INIC? Reply: INIC0 or INIC1
Trigger abort ABO Moves to the IDLE state.
Trigger source 
selection
TRS0 IMMEDIATE
TRS1 MANUAL
TRS2 EXTERNAL
TRS3 BUS
TRS4 LEVEL
TRS5 DELTA
TRS? Reply: One of TRS0 to TRS5
Item Command Description Initial value