7451A/7461A/7461P Digital Multimeter Operation Manual
6.7.3 ADC Command Reference
6-30
Calculation Comparator
calculation
LOn LOW constant
n:-9999999.E+6 to +9999999.E+6 (7461A/61P)
n:-999999.E+6 to +999999.E+6 (7451A)
Setting resolution:0.000001E-9
(0)
HIM Sets the HIGH constant to the measurement value.
LOM Sets the LOW constant to the measurement value.
HI? Reply: HId.ddddddEdd (7461A/61P) *3
Reply: HId.dddddEdd (7451A)
LO? Reply: LOd.ddddddEdd (7461A/61P) *3
Reply: LOd.dddddEdd (7451A)
PASS condition
range setting
LOP0 Calculation result LO is not specified as the PASS condition.
LOP1 Calculation result LO is specified as the PASS condition.
LOP? Reply: LOP0 or LOP1
MIP0 Calculation result GO is not specified as the PASS condition.
MIP1 Calculation result GO is specified as the PASS condition.
MIP? Reply: MIP0 or MIP1
HIP0 Calculation result HI is not specified as the PASS condition.
HIP1 Calculation result HI is specified as the PASS condition.
HIP? Reply: HIP0 or HIP1
Statistical
calculation
SIRDn, m Statistical calculation range setting and performing the calculation
n, m: 0 to 9999 (7451A/61A)
n, m: 0 to 19999 (7461P)
Note:
1. An error occurs if no data exists in the range set by the
measurement data memory.
2. No parameters can be omitted. (0, -1)
(0, -1)
SIRD? Statistical calculation range reading
Reply: SIRDdddd,dddd (7451A/61A)
Reply:
SIRDddddd,
ddddd (7461P)
Initial state: SIRD+0000, -0001 (7451A/61A)
Initial state: SIRD+00000, -00001 (7461P)
Statistical calculation result reading
SCNT? Number of samples reading
Reply: SCNT+d.ddddddE+dd (7461A/61P) *3
Reply: SCNT+d.dddddE+dd (7451A)
SMAX? Maximum value reading
Reply: SMAXd.ddddddEdd (7461A/61P) *3
Reply: SMAXd.dddddEdd (7451A)
SMIN? Minimum value reading
Reply: SMINd.ddddddEdd (7461A/61P) *3
Reply: SMINd.dddddEdd (7451A)
*3: The position of the decimal point in the reply is fixed.
Item Command Description
Initial
value