EasyManua.ls Logo

ADCMT 7461A - Page 143

Default Icon
212 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7451A/7461A/7461P Digital Multimeter Operation Manual
6.7.4 SCPI Command Reference
6-43
4. System relevant command
Function description SCPI Command Parameter Query reply
Buzzer Setting :SYSTem:BEEPer:STATe {ON,1|OFF,0} {1|0}
Error queue reading :SYSTem:ERRor? <int_3>,<str>
SCPI version reading :SYSTem:VERSion? <str>
Block delimiter :SYSTem:GPIB:DELimiter:BLOCk {CRLF|LF|EOI} {CRLF|LF|EOI}
Power supply frequency [:SENSe]:LFRQency? 50|60
Invalid state of the function read-
ing
:INHibit:FUNCtion:<Function>? {0|1}
(0: Selectable, 1: Non-
selectable)
Input terminal setting
(Front/Rear)
:ROUTe:TERMinals {FRONt|REAR} {FRON|REAR}
Measurement data display setting :DISPlay {ON,1|OFF,0} {1|0}
Measurement data memory :TRACe:STATe
Store setting {ON,1|OFF,0} {1|0}
Initialization :TRACe:CLEar
Readout range setting :TRACe:NUMBer [0|0|9999], [0|0|9999]
(7451A/61A)
[0|0|19999], [0|0|19999]
(7461P)
Data reading :TRACe:DATA? Refer to Section 6.7.2,
“Output Data Format.”
Data range reading :TRACe:DATA:NUMBer? <int_4>,<int_4>
(7451A/61A)
<int_5>,<int_5>
(7461P)
Number of measurement
results reading
:TRACe:DATA:POINts? <int_5>
Complete signal output mode :TRIGger:COMPlete SINGle|MULTi SING|MULT
Complete signal pulse width selec-
tion
:TRIGger:COMPlete:WIDTh [0.000005|0.0001|
0.0001]
<real_6>
Selection of the old model name
for the response to *IDN?
:OIDentity
This command is applied to software revision C00 or
later, and is not available on the 7461P.
{ON,1|OFF,0}
(Initial value is OFF)
{1|0}

Table of Contents

Related product manuals