7451A/7461A/7461P Digital Multimeter Operation Manual
6.7.4 SCPI Command Reference
6-43
4. System relevant command
Function description SCPI Command Parameter Query reply
Buzzer Setting :SYSTem:BEEPer:STATe {ON,1|OFF,0} {1|0}
Error queue reading :SYSTem:ERRor? <int_3>,<str>
SCPI version reading :SYSTem:VERSion? <str>
Block delimiter :SYSTem:GPIB:DELimiter:BLOCk {CRLF|LF|EOI} {CRLF|LF|EOI}
Power supply frequency [:SENSe]:LFRQency? 50|60
Invalid state of the function read-
ing
:INHibit:FUNCtion:<Function>? {0|1}
(0: Selectable, 1: Non-
selectable)
Input terminal setting
(Front/Rear)
:ROUTe:TERMinals {FRONt|REAR} {FRON|REAR}
Measurement data display setting :DISPlay {ON,1|OFF,0} {1|0}
Measurement data memory :TRACe:STATe
Store setting {ON,1|OFF,0} {1|0}
Initialization :TRACe:CLEar
Readout range setting :TRACe:NUMBer [0|0|9999], [0|0|9999]
(7451A/61A)
[0|0|19999], [0|0|19999]
(7461P)
Data reading :TRACe:DATA? Refer to Section 6.7.2,
“Output Data Format.”
Data range reading :TRACe:DATA:NUMBer? <int_4>,<int_4>
(7451A/61A)
<int_5>,<int_5>
(7461P)
Number of measurement
results reading
:TRACe:DATA:POINts? <int_5>
Complete signal output mode :TRIGger:COMPlete SINGle|MULTi SING|MULT
Complete signal pulse width selec-
tion
:TRIGger:COMPlete:WIDTh [0.000005|0.0001|
0.0001]
<real_6>
Selection of the old model name
for the response to *IDN?
:OIDentity
This command is applied to software revision C00 or
later, and is not available on the 7461P.
{ON,1|OFF,0}
(Initial value is OFF)
{1|0}