EasyManuals Logo

ADCMT 7461A User Manual

Default Icon
212 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #180 background imageLoading...
Page #180 background image
7451A/7461A/7461P Digital Multimeter Operation Manual
7.1.10 Measurement Time and Display Digits
7-18
Integration Time (IT) and Sampling Interval (SI)
Integration Time and Display Digits
7461A/61P 7451A
Integration time (IT) Integration time (IT)
Setting range Resolution Setting range Resolution
10 s to 199 s1 s100 s to 199 s1 s
200 s to 9.99 ms 10 s200 s to 9.99 ms 10 s
10 ms to 5999.9 ms 100 s 10 ms to 5999.9 ms 100 s
6 s to 10 s 200 s 6 s to 10 s 200 s
0.02 PLC to 100 PLC 0.01 PLC 0.02 PLC to 100 PLC 0.01 PLC
Sampling interval (SI) Sampling interval (SI)
Setting range Setting range
50 s to 3600 s 200 s to 3600 s
Function
DCV/DCI
DCV-Bch
2W/4W
LP-2W/4W
ACV/ACI
Diode and Continuity
ACV(AC+DC)
ACI(AC+DC)
Integration time (IT)
7461A/61P 7451A 7461A/61P 7451A
10 s (IT) 200 s 3 1/2-digit display 3 1/2-digit display 3 1/2-digit display
200 s < (IT) 2 ms 4 1/2-digit display 4 1/2-digit display 3 1/2-digit display
2 ms < (IT) 15 ms 5 1/2-digit display 5 1/2-digit display 4 1/2-digit display
15 ms < (IT) 10 s 6 1/2-digit display 5 1/2-digit display 4 1/2-digit display

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADCMT 7461A and is the answer not in the manual?

ADCMT 7461A Specifications

General IconGeneral
BrandADCMT
Model7461A
CategoryMultimeter
LanguageEnglish

Related product manuals