7451A/7461A/7461P Digital Multimeter Operation Manual
8.9 System Recovery Procedure
8-16
Table 8-6 Initial Values of Setting Parameters (1 of 3)
Parameter
When
shipping
Selecting
DFLT from
6INIT in
MENU
Selecting
P.OFF from
‘PON.LD’
‘LOAD’ in 5PARAM
*RST *CLS DCL
DFLT USER setting
Function DCV DCV Saved value DCV Selecting USER0-3 DCV
Selecting Bch HI-COM HI-COM Saved value HI-COM Selecting USER0-3 HI-COM
Range Highest Highest Saved value Highest Selecting USER0-3 Highest
Auto-range ON ON Saved value ON Selecting USER0-3 ON
Number of display digits. Maximum Maximum Saved value Maximum Selecting USER0-3 Maximum
Auto-zero ON ON Saved value ON Selecting USER0-3 ON
AC filter MED MED Saved value MED Selecting USER0-3 MED
Integration time 10 PLC 10 PLC Saved value 10 PLC Selecting USER0-3
NULL calculation OFF OFF Saved value OFF Selecting USER0-3 OFF
NULL constant 0 0 Saved value 0 Selecting USER0-3 0
COMP calculation OFF OFF Saved value OFF Selecting USER0-3 OFF
HIGH value 0 0 Saved value 0 Selecting USER0-3 0
LOW value 0 0 Saved value 0 Selecting USER0-3 0
MAX/MIN calculation OFF OFF Saved value OFF Selecting USER0-3 OFF
Selecting the MAXIMUM and
MINIMUM display
NORMAL NORMAL Saved value NORMAL Selecting USER0-3 NORMAL
Smoothing calculation OFF OFF Saved value OFF Selecting USER0-3 OFF
Smoothing count 10 10 Saved value 10 Selecting USER0-3 10
Scaling calculation OFF OFF Saved value OFF Selecting USER0-3 OFF
Constant A 1 1 Saved value 1 Selecting USER0-3 1
Constant B 0 0 Saved value 0 Selecting USER0-3 0
Constant C 1 1 Saved value 1 Selecting USER0-3 1
dB calculation OFF OFF Saved value OFF Selecting USER0-3 OFF
Constant D 1 1 Saved value 1 Selecting USER0-3 1
Sampling HOLD OFF OFF Saved value OFF Selecting USER0-3 OFF
Sampling rate SLOW2 SLOW2 Saved value SLOW2 Selecting USER0-3 SLOW2
Sampling interval 500 ms 500 ms Saved value 500ms Selecting USER0-3 500 ms
Trigger delay 0 0 Saved value 0 Selecting USER0-3 0
Sampling count 1 1 Saved value 1 Selecting USER0-3 1
Level trigger
LEVEL
0 0 Saved value 0 Selecting USER0-3 0
Blank: No Action
Saved value: Last setting value when the power is turned off.