95
Product Specifications
Appendix B
Electrical:
Power Requirement: 100 to 130 VAC (60 Hz) or 200 to 240 VAC (50 Hz)
Fuses (2): 5-Amp time delay for 100-130 VAC
2.5-Amp time delay for 200-240 VAC
Power Consumption: 350 watts
Memory Backup: Integral lithium cell; 10 years typical
No power applied, not user-replaceable
Sample Volume: 2.0 to 2.5-mL
Sample Capacity: Holds up to five continuously-loadable 10-sample cassettes
plus insertable single-sample stat test
Readout: 20-character digital display and on-board thermal print-
er
Units: m°C or m°H
Range: 0 to -1000 m°C or m°H
Linearity: Typically within ±0.5% (straight line method)
Resolution: ±1 m°C or m°H
Repeatability: ± 2 m°C (1 S.D.)
Communications: DTE EIA-232/V.24 (RS-232)
Storage
Temperature: -40 to +45°C (-40 to +113°F)
Operating Temp.
(ambient): 18 to 35°C (64 to 95°F)
Drift: Less than 1 unit per month plus 1 unit for every 5°C (9°F)
ambient temperature change