At “Select Beeper”, press START to display “ [ON]< Beeper *
[OFF] ”. The current setting is indicated by the symbol, “*”.
At “ [ON]< Beeper * [OFF] ”, press < or > to toggle the keypad
beeper on or off. Neither selection will affect the invalid-selection
beeper except that, when the keypad beeper is on, an invalid selec-
tion will invoke two beeps; when off, only one.
Press ENTER to store the new setting (or STOP to restore the orig-
inal setting).
8. Set Serial Rate
This menu item enables changing the serial rate of the data sent to
the communication port.
At “Set Serial Rate”, press START to display “Serial Rate =
xxxx”.
At “Serial Rate = xxxx”, the serial-port baud rate may be reset
by means of the < or > button to either 300, 1200 or 9600 bps.
Press ENTER to store the new setting (or STOP to restore the orig-
inal setting).
9. Select Data Capture
The EIA-232/V.24 communication port normally communicates
only the final result of each test. When implemented, this menu
item enables or disables communication of data taken throughout
each freezing point curve to a PC with data capture software for
troubleshooting or research purposes. Data Capture transmits data
at 9600 bps and sets the communication port serial rate accordingly.
Do not select any other serial rate while Data Capture is enabled.
At “Select Data Capture”, press START to display “ [ON]<
Capture *[OFF] ”. The current setting is indicated by the symbol,
“*”.
25