Chapter 3 Troubleshooting
26
Anoxomat
®
III Service Guide
3.11 Gas connection errors
Error Possible cause/Suggested action
Gas
connection
has not
passed the
corresponding
test
Cause: No pressure from gas cylinder.
Action: Check if gas cylinder is open
Cause: Pressure supplied to
Anoxomat should be 1.75 ± 0.15 bar
(25.5 ± 2.0 psi)
Action: Check output pressure on the
regulator
Cause: The regulator gauge does not
indicate actual pressure
Action: Check actual pressure in the
Process Report and adapt the setting
of the regulator
Cause: Gas cylinder is almost empty
and can no longer provide stable
pressure.
Cause: Check cylinder pressure on
the regulator
Action: Gas flow is restricted (e.g., a
kink in the gas tube) Check the gas
tube.
3.12 Pump time exceeded error
Error Possible cause/Suggested action
Pump time exceeded.
Cause: Jar was disconnected during
evacuation.
Action: Reconnect the jar
Cause: Lid of jar was opened during
recipe.
Action: Close and retighten the clamp
Cause: Pump is losing capacity.
Action: Pump needs to be cleaned or
serviced..
3.13 Fill time exceeded errors
Error Possible cause/Suggested action
Fill time exceeded.
Cause: Enough pressure but not
enough gas flow.
Action: Check for kinks in the gas
tube. Open the gas supply fully or
replace gas cylinder if it is nearly
empty
Cause: Gas inlet port restricted or
blocked by debris
Action: Contact your local Anoxomat
distributor or Advanced Instruments
Cause: Jar was disconnected during
filling.
Action: Reconnect jar and restart the
recipe
Cause: Lid of jar was opened during
program.
Action: Close and retighten the
clamp. Restart the recipe
Cause: Lid was not tightened enough
with the jar lid clamp, allowing gas to
escape during last stages of the filling
process.
Tighten the jar clamp and restart.

Table of Contents

Other manuals for Advanced Instruments Anoxomat III

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments Anoxomat III and is the answer not in the manual?

Advanced Instruments Anoxomat III Specifications

General IconGeneral
BrandAdvanced Instruments
ModelAnoxomat III
CategoryMedical Equipment
LanguageEnglish

Related product manuals