To perform a system update, follow these steps:
Insert a USB memory drive with valid update files.
Select an update (Display Board or Control Board).
Press START.
Once the update process is completed, power off the instrument, then re-power.
Verify the new revision is printed upon start up, if the printer is enabled.
Press the information icon and go to Revisions to confirm the revision shown.
7.9.2 Databases
To clear or export the various databases, select the Databases tab. (Figure 7.26). Here the user
can select which database to export or clear.
To export a table, select the desired table: Results, Events, Users, or Controls. Insert a
USB memory drive and press EXPORT .CSV to export the database.
To delete a table, select the desired table. Press CLEAR.
The Raw Database can also be exported. To export the raw data, select the desired
information (Results or Configuration). Press EXPORT .DB.
Press EXIT to return to the Settings screen.
Figure 7.26: Service (Databases)