OsmoPRO® Multi-Sample Micro-Osmometer User’s Guide
78
C - Known limitations
Additives such as anticoagulants from collection tubes may contribute to the measured
osmolality.
Particulate matter can cause premature crystallization, i.e. sample pre-freeze. The
particulate matter may be eliminated in blood and urine by centrifugation. It may also be
eliminated in urine by filtration.
Samples with air bubbles do not reliably freeze and may yield inaccurate results.
Samples allowed to evaporate will yield higher osmolality values.
Non-aqueous solutions will not yield accurate results. The OsmoPRO is intended to test
aqueous solutions only.
Highly viscous solutions may not reliably freeze and may yield inaccurate results.
High concentrations of ethanol, acetone, methanol, isopropanol, ethylene glycol, diethyl
ether, paraldehyde, trichloroethane, or propylene glycol may not reliably freeze and may
yield inaccurate results.