OIW EX SERIES-SIDE STREAM
OPERATION MANUAL
Rev. 004 MAY 2015
A3.1.3 – Issues Addressed
Erroneous log messages created during operation, boot up or upgrade which were misleading and/or incorrect
are now eliminated
In previous V5 releases a dual line system would not trigger a dry contact in a PPM over threshold situation for
Water line 2. Fixed.
In previous V5 releases related to Dual probe systems, both probes would “auto tune” but the frequency
allocated to Probe 1 would be used for both probes during cleaning. Fixed.
In previous V5 releases an independent calibration for each water line on an EX-100 Dual Stream was not
possible. Fixed.
The communication between the software and the spectrometer meant on occasion the system would lose
readings for a short period of time. Fixed.
Some internal code optimisation also carried out.