EasyManua.ls Logo

AEMC 205 - Button

AEMC 205
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
2.6 BUTTON
This button is used to display and store the reference value in the unit of
magnitude measured, or to display the differential and relative values, in %.
NOTE: This button is not functional in phase rotation mode.
Successive
presses on
Function
short
Enters the ΔREL mode, to store, then display the reference
value. The ΔRef symbol is displayed.
- Displays the differential value:
(current value – reference (∆))
The ΔREL symbol is displayed.
- Displays the relative value in %
(current value – reference (∆))
The ΔREL and % symbols are displayed.
- Displays the reference. The ΔRef symbol is displayed
- Displays the current value. The ΔRef symbol blinks.
long (>2 sec)
- Exits from the ΔREL mode
NOTE: The “Relative mode
ΔREL” function can also be used with the functions of
the MAX/MIN/PEAK mode.

Table of Contents