ACCEPTANCE TESTING
5-16
(11) Refer to Results table 5-18 on page 5-42. Set the UUT Mod Gen 2 to provide 10 kHz
deviation level at a frequency of 50 Hz.
(12) Select the 30 Hz to 50 kHz filter on the Modulation meter.
(13) Check that the Modulation meter reads 10 kHz deviation ±10% ±10 Hz. Repeat for UUT
modulation frequencies of 10 kHz and 15 kHz.