ACCEPTANCE TESTING
5-34
RF broad band power meter
Specification
Frequency range:
200 kHz to 1.05 GHz.
Accuracy:
± 10% ± resolution.
Resolution:
0.1 dB
Test equipment
Description Minimum specification Example
Calibrated RF
power source
consisting of:-
Accuracy better than 2.1%.. See below.
RF signal
generator
11 MHz to 1 GHz frequency range, RF level
13 dBm
IFR 2041
Power splitter
6 dB 50 Ω 11 MHz to 1 GHz.
Weinschel 1870A
RF Amplifier 3W 40 dB gain 11 MHz to 1000 MHz IFR 2177 or
AR5W1000 MHz
Attenuator pads Values dependent upon amplifier used.
RF power meter
±0.1 dB from 11 MHz to 1 GHz
IFR 6960/A/B +
6912 sensor
(1) Refer to Results table 5-45 on page 5-51. Set the UUT to [Tx TEST], [SELECT] N-type
input socket (i.e. only the K LED on). From power up, the instrument should already be set
to Autotune mode with wide band power meter (WB) selected.
(2) Connect the equipment as shown in Fig 5-19, connecting to the N-type input socket on the
UUT.
(3) Set the calibrated power source to provide a signal at 11 MHz and 100 mW(+20 dBm) to the
UUT input. Note the level on the UUT broad band power meter and check that it is within
the stated specification. Repeat at 100 MHz and then in 100 MHz steps up to and including
1000 MHz.
Power meter
Splitter
Pad
Pad Pad
RF amplifier
Power
sensor
Signal generator
C6054
RF OUTPUT
SENSOR
INPUT
N TYPE
INPUT
UUT
Fig. 5-19 RF broad band power meter checks