EasyManuals Logo

Agilent Technologies 1260 Infinity G1315C User Manual

Agilent Technologies 1260 Infinity G1315C
314 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #114 background imageLoading...
Page #114 background image
114 Agilent 1260 Infinity DAD and MWD User Manual
5
How to optimize the Detector
Optimizing Selectivity
With a UV-visible detector based on a diode array and the correct choice of a
reference wavelength setting, quantitative detection is possible. To suppress
caffeine, the reference wavelength must be set to 282 nm. At this wavelength,
caffeine shows exactly the same absorbance as at 222 nm. When the
absorbance values are subtracted from each another, any indication of the
presence of caffeine is eliminated. In the same way, hydrochlorothiazide can
be suppressed if caffeine is to be quantified. In this case the wavelength is set
to 204 nm and the reference wavelength to 260 nm. Figure 41 on page 114
shows the chromatographic results of the peak suppression technique.
The trade-off for this procedure is a loss in sensitivity. The sample signal
decreases by the absorbance at the reference wavelength relative to the signal
wavelength. Sensitivity may be decreased by as much as 10–30 %.
Figure 41 Peak Suppression Using Reference Wavelength
7LPHPLQ 7LPHPLQ
:DYHOHQJWKQP
1RUHIHUHQFH
+\GURFKORURWKLD]LGH
DQGFDIIHLQH
:DYHOHQJWKQP
1RUHIHUHQFH
+\GURFKORURWKLD]LGH
DQGFDIIHLQH
+\GURFKORURWKLD]LGHV
XSSUHVVHG
:DYHOHQJWKQP
5HIHUHQFHQP
&DIIHLQHVXSSUHVVHG
:DYHOHQJWKQP
5HIHUHQFHQP

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity G1315C and is the answer not in the manual?

Agilent Technologies 1260 Infinity G1315C Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity G1315C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals