Chapter 3: Testing Logic Analyzer Performance
Test Pods 3 and 4 in 600 Mb/s Mode
the 600 Mb/s mode because the other clocks are not available in this mode.
Upon completion, the logic analyzer is completely tested.
6 Complete the Performance Test Record on page 60.
Conclude the State Mode Tests
Do the following steps to properly shut down the logic analyzer session after
completing the state mode tests.
1 End the test.
a From the Main Menu, choose FileExit. At the dialog “Do you want to
save the current configuration?” select No.
Ending and restarting the logic analysis session will re-initialize the system.
b Disconnect all cables and adapters from the pulse generator and the
oscilloscope.