Chapter 3: Testing Logic Analyzer Performance
Testing the module using a 16700-series Mainframe
1 Using the mouse, drag the sample position (blue line) of a stray channel
(channel Label1 [2] in the above example) so that it is in the same eye as
the other channels. The Suggested Position from Eye Finder (green
triangle) will also move to the new eye.
2 Repeat the above step for all remaining stray channels.
3 Select Run Eye Finder. Eye Finder will recalculate the new sample
positions based on the sample position changes. The following example
shows all sampling positions aligned and in the correct eye.