Chapter 5 Error Messages
Execution Errors
5
-221 Settings conflict;
trigger output connector used by burst gate
If you have selected the gated burst mode (BURS:MODE GAT command)
with burst enabled, the “trigger out” signal cannot be enabled
(OUTP:TRIG ON command). The rear-panel Trig connector cannot be
used for both operations at the same time.
-221 Settings conflict;
trigger output connector used by trigger external
When the
external
trigger source is selected (
TRIG:SOUR EXT
command
),
the function generator automatically disables the “trigger out” signal.
The rear-
panel
Trig
connector cannot be used for both operations at the
same time
.
-221 Settings conflict;
frequency reduced for user function
For arbitrary waveforms, the output frequency is limited to 25 MHz.
When you change to the arbitrary waveform function (APPL:USER or
FUNC:USER
command) from a function that allows a higher frequency,
the function generator will automatically adjust the frequency to 25 MHz.
-221 Settings conflict;
frequency reduced for pulse function
For pulse waveforms, the output frequency is limited to 50 MHz.
When you change to the pulse function (APPL:PULS or FUNC:PULS
command) from a function that allows a higher frequency, the function
generator will automatically adjust the frequency to 50 MHz.
-221 Settings conflict;
frequency reduced for ramp function
For ramp waveforms, the output frequency is limited to 1 MHz.
When you change to the ramp function (APPL:RAMP or FUNC:RAMP
command) from a function that allows a higher frequency, the function
generator will automatically adjust the frequency to 1 MHz.