255
Chapter 5 Error Messages
Execution Errors
4
5
-221 Settings conflict;
frequency made compatible with burst mode
For an internally-triggered burst, the output frequency is limited to a
minimum of 2 mHz. The function generator has adjusted the frequency
to be compatible with the current settings.
-221 Settings conflict;
frequency made compatible with FM
When FM is enabled, the output frequency of the carrier waveform is
limited to a minimum of 5 Hz. The function generator has adjusted the
frequency to be compatible with the current settings.
-221 Settings conflict;
burst turned off by selection of other mode or modulation
The function generator will allow only one modulation, sweep, or burst
mode to be enabled at the same time. When you enable a modulation,
sweep, or burst mode, all other modes are turned off.
-221 Settings conflict;
FSK turned off by selection of other mode or modulation
The function generator will allow only one modulation, sweep, or burst
mode to be enabled at the same time. When you enable a modulation,
sweep, or burst mode, all other modes are turned off.
-221 Settings conflict;
FM turned off by selection of other mode or modulation
The function generator will allow only one modulation, sweep, or burst
mode to be enabled at the same time. When you enable a modulation,
sweep, or burst mode, all other modes are turned off.
-221 Settings conflict;
AM turned off by selection of other mode or modulation
The function generator will allow only one modulation, sweep, or burst
mode to be enabled at the same time. When you enable a modulation,
sweep, or burst mode, all other modes are turned off.