Chapter 5 Error Messages
Execution Errors
5
-221 Settings conflict;
offset changed due to amplitude
The relationship between offset voltage and
output amplitude is shown
below. Vmax is the maximum peak voltage for the selected output
termination (5 volts for a 50 load or 10 volts for a high-impedance load).
If the specified offset voltage is not valid, the function generator will
automatically adjust it to the maximum dc voltage allowed with the
amplitude specified.
|Voffset| <
Vmax
-221 Settings conflict;
amplitude changed due to offset
The relationship between output amplitude
and offset voltage is shown
below. Vmax is the maximum peak voltage for the selected output
termination (5 volts for a 50 load or 10 volts for a high-impedance load).
If the specified amplitude not valid, the function generator will
automatically adjust it to the maximum value allowed with the offset
voltage specified.
Vpp <
2 X (Vmax |Voffset|)
-221 Settings conflict;
low level changed due to high level
You can set the levels to a positive or negative value, but note that the
high level must always be greater than the low level. If you specify a high
level that is less than the low level, the function generator will
automatically set the low level to be 1 mV less than the high level.
-221 Settings conflict;
high level changed due to low level
You can set the levels to a positive or negative value, but note that the
high level must always be greater than the low level. If you specify a low
level that is greater than the high level, the function generator will
automatically set the high level to be 1 mV greater than the low level.
Vpp
2
-----------