Chapter 5 Error Messages
Execution Errors
4
5
-222 Data out of range;
burst period limited by length of burst; value clipped to upper limit
It is not possible to specify a burst period which is too short for the
function generator to output with the specified burst count and
frequency (see below). If the burst period is too short, the function
generator will automatically adjust it as needed to continuously
re-trigger the burst.
Burst Period > + 200 ns
-222 Data out of range;
burst count limited by length of burst; value clipped to lower limit
If the
Immediate
trigger source is selected (
TRIG:SOUR IMM
command
),
the burst count must be less than the product of the burst period and the
waveform frequency as shown below.
Burst Count < Burst Period X Waveform Frequency
-222 Data out of range;
amplitude; value clipped to ...
This generic message indicates that the waveform amplitude has been
limited to an upper or lower boundary.
-222 Data out of range;
offset; value clipped to ...
This generic message indicates that the offset voltage has been limited to
an upper or lower boundary.
-222 Data out of range;
frequency in burst mode; value clipped to ...
This generic message indicates that the frequency has been limited to an
upper or lower boundary as dictated by the burst period.
-222 Data out of range;
frequency in FM; value clipped to ...
This generic message indicates that the carrier frequency has been
limited to a lower boundary as determined by the FM:DEV command.
The carrier frequency must always be greater than or equal to the
frequency deviation.
Burst Count
Waveform Frequency
-----------------------------------------------------------

Table of Contents

Other manuals for Agilent Technologies 33250A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33250A and is the answer not in the manual?

Agilent Technologies 33250A Specifications

General IconGeneral
BrandAgilent Technologies
Model33250A
CategoryInverter
LanguageEnglish

Related product manuals